|
|
||||||||
a Dep. of Physical Geography and Soil Sci., Univ. of Amsterdam, Nieuwe Prinsengracht 130, 1018 VZ Amsterdam, The Netherlands
s.huisman{at}frw.uva.nl
| ABSTRACT |
|---|
|
|
|---|
Abbreviations: TDR, time domain reflectometry
| INTRODUCTION |
|---|
|
|
|---|
soil). Giese and Tiemann (1975) suggested to calculate the load resistance RL (
) from the reflection coefficient at long times (
). To avoid overestimation of the sample resistance, RL should be corrected for resistance of the cable per meter Rc (
m-1) and the combined extra series resistance R0 (
) caused by connectors and cable tester (Heimovaara et al., 1995):
![]() | (1) |
Heimovaara et al. (1995) and Mallants et al. (1996) calibrated cable and probe properties (R0, Rc and Kp) using a least-squares fitting of TDR conductivity measurements in solutions of different concentrations to conductivity measurements made with a laboratory conductivity meter (Radiometer Copenhagen CDM83 conductivity meter with CDC314 measurement cell). Recently, Reece (1998) proposed a method to measure Rc, R0, and Kp directly. He suggested that direct measurements would eliminate the need for a time-consuming calibration in multiple solutions. His method is based on resistance measurements of a coaxial cable with and without a shorted transmission waveguide. For his two-wire probes, direct measurement of Rc, R0, and Kp resulted in reliable conductivity measurements, but the cable and probe properties obtained with least-squares fitting resulted in even more accurate conductivity measurements. He stated that the higher accuracy obtained with the least-squares fitting method indicated that the fitting procedure adjusts Rc, R0, and Kp to account for small measurement errors that are not necessarily associated with these parameters (overfitting). However, the higher accuracy may also suggest that the theory is incomplete and that the fitting procedure corrects for the deviations from the theory. Besides the comparison of calibration and direct measurements, this suggestion of overfitting from Reece (1998) is also examined in this note.
| Methods |
|---|
|
|
|---|
![]() | (2) |
![]() | (3) |
is the permitivity of the material in the probe (-), c is the speed of light (3 x 108 m s-1), L is the length of the probe (m), Z0 is the characteristic impedance of the probe (
),
0 is the dielectric permitivity of free space (8.854 x 10-12 F m-1), a is the outer diameter of the inner wire, and b is the inner diameter of the outer wires (m). Alternatively, Z0 can be inferred from measurements of the reflection coefficient (
t) at a time t, which corresponds to the location of the waveguide in a nonconductive medium of known permitivity,
, according to
![]() | (4) |
). In a lossy or dispersive medium the reflection coefficient will be diminished by conduction, thus reducing the ratio in Eq.[4]. Generally, it is suggested to measure
t in a lossless medium like demineralized water, but the choice of the position of
t remains arbitrary (Heimovaara, 1992; Baker and Spaans, 1993). To compare calibration with direct measurement (Reece, 1998), a total of 25 measurements were made in five KCl solutions with conductivities of 0, 11, 119, 232, and 439 mS m-1 for five cable lengths (ranging from 5 to 25 m of RG58 C/U coaxial cable). We used a three-wire probe (Heimovaara, 1993) and a Tektronix 1502C cable tester (Beaverton, OR). The methods to obtain Rc, R0, and Kp directly [Kp with Eq. [2] and [3] (Method 1) or Eq. [2] and [4] (Method 2)] are compared with results of least-squares fitting of Kp with directly measured cable properties (Method 3), and with least-squares fitting of both Kp and cable properties (Method 4). Measurements were randomly separated into a calibration set (12 measurements) and a validation set (13). Subsequent calibration and validation were repeated 150 times for different sets.
| Results and Discussion |
|---|
|
|
|---|
m-1 and an internal cable tester and connector resistance of 0.080
(Fig. 1)
. Resistance measurements with shorted probes attached to the cables resulted in an upward resistance shift, which resulted in a probe resistance of 0.373
and a total internal resistance R0 of 0.453
. This confirms Reece's (1998) speculation that the probes used by Heimovaara et al. (1995) had a relatively high R0.
|
|
A reduction of calibration time by direct measurement of Rc, R0, and Kp does not appear feasible in the case of three-wire probes. A simplification of the calibration procedure might be considered when extensive calibration in multiple solutions is a problem. One possibility is to reduce the fitting procedure to only two cable lengths and two solution concentrations. For the validation set this approach resulted in a low root mean square error of 1.152 when cables of 25.6 and 4.1 m and solutions of 120 and 439 mS m-1 were used. Other calibration sets of four measurements also resulted in accurate conductivity measurements, provided that the selected cable lengths and solution concentrations covered the range of expected values. This illustrates that only four well chosen measurements contain all the information needed to obtain Rc, R0, and Kp, as can be expected from the simple nature of the underlying equations.
In summary, although the methods presented by Reece (1998) do have the advantage of being direct measurements, care should be taken in the application of these methods to three-wire probes. For a three-wire probe system, least-squares fitting of probe and cable properties resulted in the most accurate conductivity measurements. Comparison of several calibration techniques suggested that directly measured cable and probe properties deviate from their calibrated (optimal) values. It is certainly important (and curious) to note that there are unexplained differences between direct and empirical methods. Furthermore, it was suggested that a reduction of calibration time can be achieved by limiting the calibration procedure to two well chosen cable lengths and solution concentrations.
| ACKNOWLEDGMENTS |
|---|
Received for publication November 2, 1998.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C.-P. Lin, C.-C. Chung, J. A. Huisman, and S.-H. Tang Clarification and Calibration of Reflection Coefficient for Electrical Conductivity Measurement by Time Domain Reflectometry Soil Sci. Soc. Am. J., June 18, 2008; 72(4): 1033 - 1040. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Huisman, C. P. Lin, L. Weihermuller, and H. Vereecken Accuracy of Bulk Electrical Conductivity Measurements with Time Domain Reflectometry Vadose Zone J., April 14, 2008; 7(2): 426 - 433. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-P. Lin, C.-C. Chung, and S.-H. Tang Accurate Time Domain Reflectometry Measurement of Electrical Conductivity Accounting for Cable Resistance and Recording Time Soil Sci. Soc. Am. J., June 29, 2007; 71(4): 1278 - 1287. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Castiglione and P. J. Shouse The Effect of Ohmic Cable Losses on Time-Domain Reflectometry Measurements of Electrical Conductivity Soil Sci. Soc. Am. J., March 1, 2003; 67(2): 414 - 424. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Sarkar, M.E. Essington, and K.C. Misra Adsorption of Mercury(II) by Kaolinite Soil Sci. Soc. Am. J., November 1, 2000; 64(6): 1968 - 1975. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| The SCI Journals | Agronomy Journal | Crop Science | |||
| Vadose Zone Journal | Journal of Plant Registrations | ||||
| Journal of Natural Resources and Life Sciences Education |
Journal of Environmental Quality |
||||